B-DETECTION PLUS m – mobile gas measuring system for high-pressure applications
B-DETECTION PLUS m is a compact, portable all-in-one solution that gives you the freedom to perform reliable breathing air measurement wherever and whenever you need to.
Like the stationary models, the B-DETECTION PLUS m delivers completely reliable, ultra-precise readings to monitor compliance with DIN EN 12021:2014 limits for CO, CO2, O2 and – optionally – absolute humidity and total oil values (VOC)1.
B-DETECTION PLUS m offers an array of measurement options. The gas sampling unit supplied as standard is used to measure air quality in breathing air cylinders. Optionally, measurement can also be performed directly at the compressor. As a further option, intake air can be measured before it enters the compressor with the optional ambient air pump.
The control unit allows a tailored measurement profile to be selected for measuring air directly from the compressor or air cylinder. If limit values are exceeded, the B-DETECTION PLUS m control unit activates visual alarm signals. Solid legal certainty is ensured by an integrated data logger with SD card function for performing defined individual measurements. These can be read out and displayed separately via an BAUER Excel macro.
The transport case offers IP65-rated protection against dust and water jet ingress and withstands harsh environmental conditions.
Technical data
B-DETECTION PLUS m online breathing air and gas measuring systems | |
MEASUREMENT VALUES | |
Carbon dioxide (CO2), carbon monoxide (CO), oxygen (O2), absolute humidity1, VOC1 | |
SYSTEM LIMITATIONS | |
Maximum system pressure | 350 bar |
Maximum permissible charging rate compressor | 850 l/min |
Permissible operating temperature | +5°C… +45°C |
MEASUREMENT METHOD | |
CO2 | Nondispersive infrared spectroscopy (NDIR) |
CO, O2 | electrochemical |
Absolute humidity | Via dew point |
VOC (oil) | Photoionization detector (PID) |
MEASUREMENT VALUES AND MEASUREMENT ACCURACY 3 | |
CO2 | 0… 2.000 ppm ± 2.5 % Full scale |
CO | 0 … 25 ppm ± 2 % of measurement range final value |
O2 | 0 % … 40 % O2 ± 1 % of measurement range final value |
Dew-point | -70 …+60°C / 10 … 40,000 ppm ± 2°C / 1 ppm + 20% of displayed value |
VOC (oil) | 0 … 1 ppm (based on isobutene as calibration gas) |
FURTHER TECHNICAL SPECIFICATIONS | |
Operating voltage | integrated: 24 VDC; stand-alone: 100-250 VAC, 50/60 Hz |
Flow rate | 1.0 … 3.0 l/min |
Weight | 8.9 kg |
Dimensions W × H × D [mm] | 464 x 366 x 180 |
SYSTEM TEST/CALIBRATION | |
Visual inspection and leak test | Monthly |
Functions check 2 | Min. 3-monthly |
System check | Annually |
Calibration | dew point sensor: every 24 months |